User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Daytronic 3570:Strain Gage Indicator DATE: 21-May-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 17 NUMBER OF LINES: 155 CONFIGURATION: Fluke 5700A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ S X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP Connect the UUT +SIG and -SIG to the Fluke 5700A 1.004 DISP OUTPUT HI and LO connectors. 1.005 DISP Turn the UUT on and allow to warm up for 15 minutes. 1.006 DISP On the UUT press "SET UP" twice and then "RANGE" key. 1.007 DISP Press the "ENTER" key. The display will now 1.007 DISP show the current mV/V range. 1.008 DISP Change the display to read "3. 0" mV/V. 1.009 DISP Use the following method whenever you are asked 1.009 DISP to vary the digital display. 1.010 DISP The display is varied by pressing either the upper 1.010 DISP or lower segment of the digit to be varied. 1.010 DISP The display will not wrap around. You can only vary the 1.010 DISP display from 0 up to 9, or from 9 back down to 0. 1.011 DISP Now press the "ENTER" key and the "SET UP" key twice. 1.011 DISP Wait for the display to return to its normal 1.011 DISP operating mode. 1.012 DISP On the UUT press the SET UP twice then CAL button. 1.012 DISP The UUT should display "2 Pt". 1.013 DISP Now press the ENTER key. 1.013 DISP The UUT should now display "oFSEt". 1.014 5700 0mV S 2W 1.015 DISP Press ENTER again. The UUT should now display 1.015 DISP the input from the 5700. 1.016 DISP Now press any key except CAL or ENTER. 1.017 DISP This will freeze the display and allow you to 1.017 DISP change the display to read zero. 1.018 DISP Press ENTER. The display will read "oFSEt". 1.019 5700 45mV S 2W 1.020 DISP Press "CAL". The display will now read "ForcE". 1.021 DISP Press "ENTER" and then "CAL". The display should 1.021 DISP read "SHn". Now press "CAL". 1.022 DISP Again press any key except "CAL" or "ENTER". 1.023 DISP Adjust the display to read 32700. 1.024 DISP Press "ENTER". The UUT should display "ForcE". 1.024 DISP Press "SET UP" twice and wait for the display 1.024 DISP to return to its normal operating mode. 1.025 HEAD {} 1.026 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.027 HEAD {º Linearity Checks º} 1.028 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.029 HEAD ÍÍ Linearity Checks ÍÍ 1.030 HEAD {ÍÍ 3.0mV/V ÍÍ} 1.031 JMP 2.001 1.032 EVAL 2.001 ASK- U 2.002 5700 40.5mV S 2W 2.003 MEMI Enter the UUT reading. Do not include the decimal point. 2.004 MATH mem1= 29430 2.005 MEME 2.006 MEMC ÿ 8U 3.001 5700 36mV S 2W 3.002 MEMI Enter the UUT reading. Do not include the decimal point. 3.003 MATH mem1= 26160 3.004 MEME 3.005 MEMC ÿ 8U 4.001 5700 31.5mV S 2W 4.002 MEMI Enter the UUT reading. Do not include the decimal point. 4.003 MATH mem1= 22890 4.004 MEME 4.005 MEMC ÿ 8U 5.001 5700 27mV S 2W 5.002 MEMI Enter the UUT reading. Do not include the decimal point. 5.003 MATH mem1= 19620 5.004 MEME 5.005 MEMC ÿ 8U 6.001 5700 22.5mV S 2W 6.002 MEMI Enter the UUT reading. Do not include the decimal point. 6.003 MATH mem1= 16350 6.004 MEME 6.005 MEMC ÿ 8U 7.001 5700 18mV S 2W 7.002 MEMI Enter the UUT reading. Do not include the decimal point. 7.003 MATH mem1= 13080 7.004 MEME 7.005 MEMC ÿ 8U 8.001 5700 13.5mV S 2W 8.002 MEMI Enter the UUT reading. Do not include the decimal point. 8.003 MATH mem1= 9810 8.004 MEME 8.005 MEMC ÿ 8U 9.001 5700 9mV S 2W 9.002 MEMI Enter the UUT reading. Do not include the decimal point. 9.003 MATH mem1= 6540 9.004 MEME 9.005 MEMC ÿ 8U 10.001 5700 4.5mV S 2W 10.002 MEMI Enter the UUT reading. Do not include the decimal point. 10.003 MATH mem1= 3270 10.004 MEME 10.005 MEMC ÿ 8U 11.001 HEAD {} 11.002 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 11.003 HEAD {º Range Checks º} 11.004 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 11.005 HEAD ÍÍ Range Checks ÍÍ 11.006 HEAD {ÍÍ 1.5mV/V ÍÍ} 11.007 JMP 12.001 11.008 EVAL 12.001 DISP On the UUT press "SET UP" twice and then "RANGE" key. 12.002 DISP Press the "ENTER" key. The display will now 12.002 DISP show the current mV/V range. 12.003 DISP Change the display to read "1. 5" mV/V. 12.004 DISP Now press the "ENTER" key and the "SET UP" key twice. 12.004 DISP Wait for the display to return to its normal 12.004 DISP operating mode. 12.005 5700 22.5mV S 2W 12.006 MEMI Enter the UUT reading. Do not include the decimal point. 12.007 MATH mem1= 32700 12.008 MEME 12.009 MEMC ÿ 8U 13.001 5700 11.25mV S 2W 13.002 MEMI Enter the UUT reading. Do not include the decimal point. 13.003 MATH mem1= 16350 13.004 MEME 13.005 MEMC ÿ 8U 14.001 HEAD {ÍÍ .75mV/V ÍÍ} 14.002 JMP 15.001 14.003 EVAL 15.001 DISP On the UUT press "SET UP" twice and then "RANGE" key. 15.002 DISP Press the "ENTER" key. The display will now 15.002 DISP show the current mV/V range. 15.003 DISP Change the display to read ".75" mV/V. 15.004 DISP Now press the "ENTER" key and the "SET UP" key twice. 15.004 DISP Wait for the display to return to its normal 15.004 DISP operating mode. 15.005 5700 11.25mV S 2W 15.006 MEMI Enter the UUT reading. Do not include the decimal point. 15.007 MATH mem1= 32700 15.008 MEME 15.009 MEMC ÿ 8U 16.001 5700 5.625mV S 2W 16.002 MEMI Enter the UUT reading. Do not include the decimal point. 16.003 MATH mem1= 16350 16.004 MEME 16.005 MEMC ÿ 8U 17.001 HEAD {} 17.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 17.002 DISP {º THIS COMPLETES THE VERIFICATION º} 17.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 17.003 END